Peak NameRet Time (min)Window (min)   Std 1      Std 2      Std 3      Std 4      Std 5   Conc Unit
Lithium5.720.500.090.270.901.802.70ppm
Sodium8.171.002.437.2924.3048.6072.90ppm
Ammonium10.321.000.270.812.705.408.10ppm
Potassium15.741.500.812.438.1016.3024.30ppm
Magnesium27.262.002.437.2924.3048.6072.90ppm
Calcium39.032.002.437.2924.3048.6072.90ppm

Peak NameRet Time (min)Window (min)   Std 1      Std 2      Std 3      Std 4      Std 5   Conc Unit
Fluoride7.750.500.090.270.901.802.70ppm
Acetate9.150.500.090.270.901.802.70ppm
Formate11.400.500.090.270.901.802.70ppm
Chloride16.700.752.437.2924.3048.6072.90ppm
Nitrite-N20.200.750.090.270.901.802.70ppm
Bromide23.750.750.090.270.901.802.70ppm
Nitrate-N26.100.750.270.812.705.408.10ppm
Sulfate35.551.002.437.2924.3048.6072.90ppm
Thiosulfate46.701.000.090.270.901.802.70ppm
Phosphate-P50.901.000.090.270.901.802.70ppm

CALIBRATION ...


Both cation and anion calibrations are based on a 6-point calibration curve consisting of 5 standard solutions and a zero blank.  Normal calibration procedure includes 3 replicates per standard and blank and occurs once every seven days when the instrument is active.  In between calibrations, the system is left on standby so it is always ready to begin analyzing samples.  The calibration data is based on NIST traceable Inorganic Ventures single element ion chromatography standards that are mixed to prepare the following standard concentrations for anions ...

Peak Name

Calibration

Type

Integration

Type

# of Pts

Rel. Std. Dev.

%

Coef. of

Determ.

C0

(offset)

C1

(slope)

C2

(curve)

FluoridePoint to PointArea5
1


AcetatePoint to PointArea5
1


FormatePoint to PointArea5
1


ChloridePoint to PointArea5
1


Nitrite-NPoint to PointArea5
1


BromidePoint to PointArea5
1


Nitrate-NPoint to PointArea5
1


SulfatePoint to PointArea5
1


ThiosulfatePoint to PointArea5
1


Phosphate-PPoint to PointArea5
1


​The standard concentrations for cations ...

Peak Name

Calibration

Type

Integration

Type

# of Pts

Rel. Std. Dev.

%

Coef. of

Determ.

C0

(offset)

C1

(slope)

C2

(curve)

LithiumPoint to PointArea5
1


SodiumPoint to PointArea5
1


AmmoniumPoint to PointArea5
1


PotassiumPoint to PointArea5
1


MagnesiumPoint to PointArea5
1


CalciumPoint to PointArea5
1












The following is a representative summary of the calibration results for the cation standards (04/18/2017) ... 3 replicates per standard (averaged).

The following is a representative summary of the calibration results for the anion standards (04/12/2017) ... 3 replicates per standard (averaged).